Page 53 - FiM_03_2016
P. 53
Figure 3 Variation of Percentage global Gamma Passing Rate using MU limit for prostate plans
Figure 4 Variation of Percentage global Gamma Passing Rate using MU limit for H&N plans
CONCLUSIONS. It was verified that decreasing plan MU increases segments sizes and dose delivery accuracy without a relevant deterioration of the dose distribution. Maximum value of plan MUs could be used to increase dose accuracy and reduce plan complexity; similarly to others optimization parameters, an adequate balance has to be found to achieve the best plan solution.
BIBLIOGRAPHY
[1]Das, Indra J., George X. Ding, and Anders Ahnesjo. "Small fields: non-equilibrium radiation dosimetry." Medical physics 35.1 (2008): 206-215.